The Delta 3300 Defibrillator/Transcutaneous Pacemaker Analyzer is a precision instrument for testing defibrillators
and external (Transcutaneous) Pacemakers. It combines the complete testing functions of a Defibrillator Tester,
Transcutaneous Pacemaker Analyzer and 12-Lead ECG / Arrhythmia Simulator into one compact, lightweight, and
easy to-use instrument.
The Delta 3300 tests all Defibrillators including AEDs. It measures both monophasic and bi-phasic waveforms. It is
compact, light weight and easy to use. The test results can be saved in the internal memory and uploaded to the PC
and printed using ProCom Print software (CD included with the unit).
Features and Benefits
- Energy and cardio version measurement
- Peak voltage and current reading
- Waveform storage and Play back
- Tests all Automated defibrillators
- Pacemaker loads from 50 to 1100 Ohms
- Pacer Amplitude, Rate, Pulse width
- Demand Sensitivity, Refractory period,
immunity and pulse Energy
- 12 Lead ECG, Arrhythmia, Performance
- Large LCD Display with Backlight
- Simple and easy to use
Materials Included
- Delta 3300 Defibrillator Analyzer
- 1 Defibrillator Paddle Adapter
- 1 Lithium Ion Battery
- 1 Li-Ion Battery Charger
- 1 Transcutaneous / AED
Interface Cable
- 1 RS-232 Interface Cable
- 1 Carry Case
- 1 ProCom Print Software
- 1 User Manual
Technical Specifications
General Information
- Dimensions:
- Depth: 10.25"
- Width: 6.25"
- Height: 2.5"
- Weight: 3.3 lbs
- Temperature Range
- Operating: +15°C to +35°C
- Storage: 0°C to +50°C
- Display: LCD with LED back-light
- Data Input/Outputs: RS-232C (1) for Computer control
- Power: 7.4 Volt Li-Ion Battery
100-230 VAC input, 850 mA DC Charge
- Housing: High impact ABS Plastic Case
Defibrillator
- Energy Output Measurement
- Maximum Energy: 800 J
- Display Resolution: Based on Auto Range
- Auto Range: 2 - 9.99 J, 10-99.9 J, 100-800 J
- Resolution: 0.01 J ( 2 -9.99), 0.1 J (10-99.9),
1 J (100-500)
- Accuracy: ±1 % of reading ± 0.1 J (2-9.9),
± 1J (10-99), ± 2 % (100-500) ± 1J
- Load Resistance: 50 ohms ±1 %; non-inductive (< 1 μH)
- Sampling Time: 20 uS
- Measurement Time Window: 60 ms
- Absolute Max Peak Voltage: 5000 V
- Absolute Max Peak Current: 100 A
- Pulse Width: 60 ms
- High Level Output
- High Range: 100:1 amplitude attenuation
- Low Range: 200:1 amplitude attenuation
- Waveform Storage/Playback: Discharge can be viewed via ECG outputs
and paddles.
- Sync Time Measurements:
- Timing Window:
Starts: 40 ms at each R wave peak
- Test Waveforms:
All Waveform simulations available
- Delay Time Accuracy: ±1 ms
- Charge Time Measurement: 30 to 60 in
steps of 15 seconds
Transcutaneous Pacemaker
- Test Load Range:
- 50 -1100 Ohms in steps of 100 Ohms
- Accuracy: 50 - 500 ohm ±1 %; 500 - 1100 ohm ±1.5 %
- Pulse Measurements:
- Amplitude: 4 to 400 mA (50 ohm load)
- Accuracy: ±1 % or ± 1 mA
- Max. amplitude: 400 mA (50 Ohm Load)
- Rate: 30 to 800 ppm
- Accuracy: ±1 % or 2 ppm
- Pulse Width: 0.5 to 50 ms
-
Accuracy: ±1 % or ±0.3 ms
- Demand Sensitivity Test
- Waveforms: QRS, Square(SQR),
Triangle(TRI), and Haversine (SSQ)
- Waveform Width: 40, 50, 60, 70, 80, 90, 100,
200, 300, 400 ms
- Pacer Rate: 30 to 120 ppm
- ECG Output
- Amplitude: User Selectable 0.1 - 5 mV
- Pacer Input (Load dependent)
- Amplitude (50 ohm): 0 - 10 mV
- Resolution (50 ohm): 25 μV
- Amplitude: (500 ohm): 0 - 100 mV
-
Resolution: (500 ohm): 1 mV
- Accuracy: ±2 %
- Immunity Test:
- Interference Signal: 50/60 Hz
- ECG signal Amplitude: 0.5 to 5.0 mV
- Refractory Period Measurement:
- Overal: 20 to 500 ms (both Pacing and Sensing)
-
Accuracy: ±5 ms
ECG Patient Simulator
- ECG General
- Lead Configuration: 12-lead simulation
RL, RA, LA, LL, V1-6
- Output Impedance: RL to All leads 1000
- Ohms, V Leads: 1000 ohms to RL
- High Level Output: 0.5 V/mV of low level
(lead II)
- Rates: 30, 40, 50, 60, 70, 80, 90, 100,
110, 120,130, 140, 150, 160, 170, 180,
190, 200, 220, 240, 260, 280, 300 &
310 BPM
- Accuracy: ±1 %
Amplitude: User Selectable 0.1 - 5 mV
(Lead II)
Accuracy: ± 2 % @ Lead II (SQR)
- Amplitude: User Selectable 0.1 - 5 mV
(Lead II)
- Accuracy: ± 2 % @ Lead II (SQR)
- Arrhythmia Selections:
- Ventricular Fibrillation Coarse (VFBC)
- Ventricular Fibrillation Fine (VFBF)
- Ventricular Tachycardia (VTAC)
- Atrial Fibrillation (AFIB)
- Performance Waveforms:
Sine, Square, Triangle:
- Rate: 0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 6, 7,
8, 9, 10, 15, 20, 25, 30, 35, 40, 45, 50,
55, 60 Hz.
- Amplitude: 0.1, 0.25, 0.50, 1, 1.5, 2.0,
2.5, 3.0, 3.5, 4.0, 4.5, 5.0 mV.
- Amplitude Accuracy: ±2 %
(Square Wave -Lead II 1.0 mV)
- Pacer Pulse:
Pulse Width: 10mS
Amplitude: 1.0 mV (Lead II)
Accuracy: ± 2 %
Warranty Information
|
|
|
|
|
|
|