These products have been discontinued. For comparable products, click here for the following:
Criticare VitalCare Vital Signs Monitor w/ DOX SpO2, NIBP, HR and Printer
Criticare VitalCare Vital Signs Monitor w/ Nellcor SpO2, NIBP, HR, Printer
Criticare VitalCare Vital Signs Monitor w/ DOX Sp02, NIBP, TurboTemp, Printer, HR
Criticare VitalCare Vital Signs Monitor w/ Nellcor Sp02, NIBP, TurboTemp, Printer, HR
Designed to travel everywhere you do, the VitalCareTM 506N3 series
patient monitor with integrated carry-handle attaches quickly and easily
to a rolling stand. Available in multiple configurations – with choice of
Criticare or Nellcor oximetry – the 506N3 patient monitors provide
maximum versatility and cost-effectiveness.
New smart battery management allows up to 100 NIBP measurements
on a single charging. With a compact design that is easily setup for spotcheck
or continuous monitoring, the 506N3 patient monitor adapts
readily to hospital, clinic and transport requirements.
Features
and Benefits
Proprietary ComfortCuff ™ NIBP
measures on inflation for faster,
more comfortable readings.
Choice of Criticare DOX™ digital
SpO2 technology or Nellcor
OxiMax™ SpO2 technology.
Audible and visual alarms are
adjustable providing instant
notification the way you need
them.
Optional FILAC FasTemp™
predictive temperature for quick,
accurate measurements.
Optional internal printer for
on-demand, automatic interval,
trends or multi-patient reports.
Technical Specifications Non-Invasive Blood Pressure
Technique:
Oscillometric measure upon inflation
Measurement Time: <40
seconds typical; 0 to 300mmHg;
standard adult cuff
Automatic
Measurement
Cycles: 1 - 60 minute; 2 - 4 hour; STAT
Inflation
Pressure Range: 30 to 300mmHg (Adult)
30 to 150mmHg (Pediatric)
20
to 150mmHg (Neonate)
Resolution: 1mmHg
Clinical Accuracy:
SP-10:2002
Transducer Accuracy: ±2mmHg or ±2% of reading, whichever
is
greater, over full range
STAT mode: 5 minutes of consecutive
readings
Pulse Oximetry
Range:
1% to 99%
Resolution: 1%
Accuracy (DOX only): 70% to 99% range:
±2%;
50% to 69% range: ±3%;
<50%: unspecified
Accuracy
(OxiMax Only): 70% to 99% range: ±2%;
<70%: unspecified
Method:
Dual wavelength LED
Audio: Tone with each detected pulse; pitch
varies with saturation
Heart
(Pulse) Rate
Range: 20 to 300bpm
Accuracy (DOX only):
Greater of ±1bpm or ±1%, of reading
Unspecified for <30 or >240
bpm
Update Time: Every second
Resolution:
1bpm
Pulse Volume: 10 levels or Off
Temperature (optional)
Type: FILAC FasTemp™
Method:
Predictive or Continuous
Channels: 1
Range: 30.0º to 43.0º C
Accuracy
Continuous Mode:±0.2°F (±0.1°C)
Accuracy Predictive Mode: ±0.2°F
(±0.1°C) for ranges 37°C to 39 °C
(Compliant with EN 12470-3 and
ASTM
E1112:98)
Resolution: 0.1º F or 0.1º C
Predictive Response Time:
Oral: 10 seconds typical;
Axial: 10 seconds typical;
Rectal: 10
seconds typical
Probes/Wells: Oral/Axial FILAC FasTemp [202026]
Rectal
FILAC FasTemp (optional) [202034]
Isolation Well, Oral, Axial
[202089]
Isolation Well, Rectal (optional) [202087]
Alarms
Indication: Audible; Visual
Volume:
10 levels or Off
Silence: Yes; 2 minutes or permanent
Trend Memory
Memory: Last 24 hours
Parameters:
SpO2, pulse rate, NIBP, temperature
Report Types: Single Patient or
Multi-patient;
Selectable formats
Displays
Type: LED
Parameters: NIBP, SpO2;
temperature, pulse rate; pulse bar
Status: Alarm Silence; Sensor; Low
Battery; External
Power; Patient Size
Text: Alphanumeric LCD
Languages:
English, French, German, Italian, Portuguese, Spanish, Dutch, Danish
System Output
Com Port: Serial, RS
232-compatible
Printer
Type:
Thermal array
Tabular Mode: 1 - 60 minute intervals; 2 - 24 hour
intervals;
Auto on BP; Auto on Alarm, Off
Mechanical
Weight: Base unit with
battery: 6.95 lb; 3.15kg
Size: Base unit: 8.7" (H) x 7.1" (W) x 6.5”
(D); 22cm (H) x 18cm (W) x 16.5cm (D)
Power Requirements
Voltage: 100 to
240 VAC, 50 to 60 Hz
Battery: Sealed lead acid; Life: 8 hours
Rechargeable:
6 hours
Environmental
Operating
Temperature: 32º to 113º F; 0º to 45º C
Storage Temperature: 23º to
122º F; -5º to 50º C
Operating Humidity: 15% to 90%, non-condensing
Classification
Type of Protection:
Class I equipment
Degree of Protection: Type CF defibrillator-proof
equipment
Protection against Ingress: IPX1 rating, Drip-Proof
Equipment
Warranty Information
Item#
Item Name
Our Price
Qty
Add
506DN3
VitalCare 506DN3 w/ DOX SpO2, NIBP, HR
(Out of Stock)
506DNT3
VitalCare 506DNT3 w/ DOX SpO2, NIBP, HR, Temp
(Out of Stock)
506DNTP3
VitalCare 506DNT3 w/ DOX SpO2, NIBP, HR, Temp, Printer
(Out of Stock)
506LN3
VitalCare 506LN3 w/ Nellcor SpO2, NIBP, HR
(Out of Stock)
506LNT3
VitalCare 506LNT3 w/ Nellcor SpO2, NIBP, HR, Temp
(Out of Stock)
506LNTP3
VitalCare 506LNTP3 w/ Nellcor SpO2, NIBP, HR, Temp, Printer
(Out of Stock)
506N3
VitalCare 506N3 w/ NIBP
(Out of Stock)
506NP3
VitalCare 506NP3 w/ NIBP, Printer
(Out of Stock)
506NT3
VitalCare 506NT3 w/ NIBP, Temp
(Out of Stock)
506NTP3
VitalCare 506NTP3 w/ NIBP, Temp, Printer
(Out of Stock)
Check the items you wish to purchase, then click