This product has been discontinued. For comparable products, click here for the following:
This stand-alone, bedside or transport capnograph incorporates sidestream technology for applications with both intubated and non-intubated patients. Waveforms and numeric values are easily viewed at all angles, and data can be stored and recalled with either on-screen trends, a printer, or personal computer.
Features and Benefits
- Multiple display options
- Serial and analog output
- High-resolution, vacuum flourescent display
- Breath-by-breath CO2 detection
- Flexibility to incorporate SpO2 and FiO2 technology
- Optional Remote Alarms
- View various data on-screen
- Download to PC
- Interface with polysomnographs
Technical Specifications
- Display: Vacuum Fluorescent
- Indicators Alarm, Alert, Charge, Silenced
- Flow Rate: 120 ml/min, +/-20
- C02:
- Measurement: Non-Dispersive IR absorption
- Measurement Range: 0-10% CO2
- Units: User-selectable: mmHg, Vol. %, kPa
- Display Range: 0-100 mmHg or 0-10.0% CO2 or 0-13.3 kPa
- Accuracy: ±2 mmHg; ±0.3%; ±0.3 kPa, or 4% of reading, whichever is greater (0-10% C02)
- Resolution: 1 mmHg; 0.1%; 0.1 kPa
- Calibration: 2 point Cal
- N2O Compensation: User-selectable 40% (default = OFF)
- Response Time: 375 msec to 90% of value
- Averaging: 4-breath average or breath by breath, user selectable
- Respiration Rate:
- Range: 0-150 bpm
- Averaging: 4-breath average
- Accuracy: ± 1 bpm
- Sp02 (Optional):
- Range: 0-100%
- Accuracy: ±2% at 70-100%; ±3% at 50-69%
- Averaging: 4, 8 or 16 beat, user-selectable
- Pulse Rate (Optional):
- Range: 30-250 bpm
- Accuracy: ±2% @ 30-250 bpm
- Averaging: 8 or 16 seconds, user-selectable
- Pulse Tone: Pitch corresponds to SpO2 value, Volume adjustable or OFF
- Fi02 (Optional):
- Range: 0-100%
- Accuracy: ±2%
- Response Time: < 10 sec. to 90%
- Drift: < 2% full scale per month
- Dimensions: 3.5" H x 10.0" W x 5.5" D (88.9 mm H x 254 mm W x 139.7 mm D)
- Weight: 4.8 lbs. (2.2 kg)
- Power Requirements:
- AC Charger: 115 V-, 60 Hz
- Battery: Nickel Cadmium battery, 6 VDC, internal rechargeable, not user replaceable, fully charged continuous - use life 2 hours approximate, maximum full charge time 4 hours
- Serial Output: RS232C: tabular trends, patient data; For Sleep Study: 10 hrs. @ 4 sec., 31 hrs. @ 12 sec.
Warranty information
|

|