The
DELTA 3000 tests all defibrillators including AEDs. It measures both
monophasic and bi-phasic waveforms. It tests all external pacemakers
including transthoracic, transvenous, and AV sequential models. It has
12 built-in user selectable test loads for transthoracic pacing testing.
The built-in ECG / Arrhythmia simulator allows performance testing of
any ECG equipment.
The DELTA 3000 is operated via easy to follow
menus and eight soft keys. Instructions and test measurements are
displayed on the large four line LCD display. A print key outputs test
measurements to the RS 232 serial port.
The DELTA 3000 is
delivered in a hard carrying case with various paddle adapters and
interface cables. It provides unsurpassed biomedical testing capabilites
in one instrument.
Features and Benefits
- Compact, Lightweight, and Rugged
- Simple, Menu Driven Operation
- Large 4-line LCD
- Simultaneous Display of All Test Measurements
- Dedicated Print Key for Immediate Documentaion
- RS 232 Interface
- Battery or AC Operation
- Battery or AC Powered Operation
Measurements Tested
- Energy, Peak Voltage, Peak Current, and Cardio Sync. Time
- Storage and Playback of output waveforms
- All External Pacemakers - Transthoracic, Transvenous, and AV Pacers
- Amplitude, Rate, Pulse Width, Pulse Energy, Sensed and Paced Refactory Periods
- 50 and 60 Hz Immunity Tests
- Transthoracic Test Loads: Fixed 50 Ohm or 12 User Selectable Loads to 1000 ohms
- A / V Pacer Test Load: 500 Ohm
- Tests all ECG Equipment
- 12 lead ECG, User Selectabel Rates and Amplitudes
- Performance Waveforms: Sine, Square, Triangle amp; Pulse
- 12 Arrhythmias
Materials Included
- Delta 3000 Defibrillator & External Pacer Analyzer
- (10) ECG Snap to Banana Adapters
- (2) Pacer Cables
- (1) Interface Cable
- (1) AC Adapter
- (1) Hard Carrying Case
- User Manual
Technical Specifications
General
- Power: 2 9V Batteries or AC Adapter
- Size: 10" x 11" x 2.25"
- Weight: 4 lbs
Defibrillator Analyzer
- Energy Measurement
- Load: 50 Ohm +/- 1%
- Range (High): 900 J; Resolution: 1 Joule
- Range (Low): 0 - 100; Resolution: 0.1 Joules
- Accuracy: +/- 2% of the reading + 1 J for Hi Range or + 0.5 J for Low Range
- Maximum Voltage: 5200 V
- Maximum Current: 100 Amp (H)
- Sync. Time: 0 - 250 ms
- Waveform Output
- Real Time: Output 1000 : 1 Stored
- Discharge waveform expanded 200:1 and is repeated 8 sec. It can be viewed through the paddles and through the ECG High Level Output.
Pacemaker Analyzer
- Transcutaneous
- Test Load: 50 Ohms fixed; and variable 100 to 1000 Ohms (user selectable)
- Current: 1 to 200 mA ; Accuracy +/- 5% + 0.5 mA
- Pulse Width: 0.5 to 80 ms Accuracy: +/- 1% + 0.1 ms
- Refractory: 20 - 500 ms Accuracy: +/- 2% or + 10 ms
- Atrial Ventricular:
- Test Load: 500 Ohms
- Current: 1 to 25 mA Accuracy: +/- 5% + 0.5 mA
- Pulse Rate: 30 to 800 ppm. Accuracy: +/- 1% + 0.1 ms
- Refactory: +/- 20 - 500 ms
- Accuracy: +/- 2% or 10 ms
- Immunity Test: 50 / 60 Hz
ECG Arrhythmia Simulator
- ECG: Normal Sinus Rhythm
- Rate: 30, 60, 90, 120, 240
- Performance Waveforms:
- Sine, Square, Triangle and Pulse
- Rate: 0.5, 1, 2, 10, 15, 20, 25, 40 Hz
- Arrhythmias: VFBC, VFBF, VTAC, AFIB, PVC, PVC1, PVC2, BGY, RBB, AFIB, ATRFT, NSR
- Waveform Outlet: 12 Lead ECG and Paddle and High Level Output Jacks
Warranty Information
|
|
|
|
|
|
|