This product has been discontinued. For a comparable product, click here for the Criticare nGenuity 8100E Patient Monitor.
Advanced surgical
techniques, new anesthetics, the need to lower healthcare
costs, they're all pointing toward more outpatient surgeries and
procedures in hospitals, surgical centers and physician offices. That's
why you need the Welch Allyn Atlas monitor, a vital signs monitor
specifically designed to monitor patients during conscious sedation and
IV anesthesia procedures for these markets.
Designed with input
from physicians and nurse anesthetists, this conscious sedation monitor
offers bright, clear,
across-the-room visibility with large LED
readouts and crisp waveform displays, giving you at-a-glance
convenience and more time to concentrate on your patient. The intuitive
design means users can learn to operate the monitor in just minutes.
The
Atlas monitor goes where your patient goes, effortlessly. Lightweight
and portable, it can be set on a cart or countertop, wheeled on its own
dedicated mobile stand, mounted on the wall or carried using the
integrated side handle.
We offer the flexibility to configure the
monitor that best fits your needs, and it�s upgradeable, should your
needs change in the future.
Best of all, it's at a cost that you
can afford.
Designed
specifically for conscious sedation and IV anesthesia in a wide variety
of healthcare sites.
Features and Benefits
- Hard-working, easy-to-use monitor with configuration options to meet your individual needs
-
Bright, crisp waveforms and multicolored LED displays for across-the-room visibility
-
Simple, straightforward menu operation
Technical
Specifications
3-Lead or
5-Lead Selectable ECG
- Heart Rate Range: 20-250 bpm.
- Heart
Rate Accuracy: ±3 bpm or ±3%, whichever is greater.
- Bandwidth:
0.5-40 Hz (monitor mode) or 0.05-100 Hz (extended mode).
- Display
Sweep Speeds: 25, 12.5 or 6.25 mm/sec.
Noninvasive Blood Pressure
- Systolic
Measurement Range: 60-250 mmHg.
- Diastolic Measurement Range:
30-160 mmHg.
- Blood Pressure Accuracy: Meets AAMI SP10-1992
Standard.
- Operating Modes: Manual (single determination) or
Automatic Mode.
- Cycle Interval Choices: 1, 3, 5, 10, 15, 30 or
60 minutes.
- Pulse Oximetry (Nellcor Puritan Bennett™)
- Measurement
Range: 40-100%
- Accuracy: ±2 digits (no motion), ±3 digits (with
motion) in the range of 0-100%
- Pulse Rate Range: 20-250 bpm.
- Pulse
Rate Accuracy: ±3 digits (no motion), ±5 digits (with motion)
Temperature
- Range: 17-50° C
(62.6-122° F)
- Accuracy: Meets ASTM Accuracy Standards.
- Probe
Type: YSI Series 400 Compatible.
Impedance Respiration
- Range: 5-100 breaths per
minute.
- Accuracy: ±3 breaths per minute.
End Tidal Carbon Dioxide
- Measurement
Technique: Sidestream.
- Measurement Range: 0-98 mmHg.
- ETCO2
Accuracy: 0-40 mmHg ±3 mmHg; 41-98 mmHg ±10% of reading.
- Breath
Rate Measurement Range: 5-100 breaths per minute.
- Breath Rate
Accuracy: ±1 breath per minute or ±5%, whichever is greater.
General
- Size: 35 cm x 23
cm x 25 cm (13.8” x 9.1” x 9.8”)
- Weight: 4.3-5.5 kg (9.5-12.1
lbs) dependent on configuration.
- Power Requirements: 120 VAC,
50/60 Hz; 240 VAC, 50/60 Hz.
- Printer: Thermal array, prints
waveforms and numerical trends.
- Battery: Lead acid,
rechargeable.
Warranty Information
DocumentationBrochure