Home > Vital Sign Monitors >

Mindray DPM4 Masimo SpO2 two IBPs, recorder
Mindray DPM4 Masimo SpO2 two IBPs, recorder


 
Additional Views:





Product Code: 8002F-PA00001
Discontinued

Details
 

This product has been discontinued. Click here to return to Vital Signs Monitors.

Replacing the PM-8000 Express, the DPM4 is a portable patient monitor ideal for use across a variety of clinical settings due to its compact size and simple design. Intended to meet basic continuous monitoring needs, it focuses on providing core capabilities in an uncomplicated manner. Its high resolution color display offers a surprisingly vibrant view of your patient's physiological data, making it practical for use at the bedside or during transport.

Available with or without invasive blood pressure, DPM4 delivers the flexibility you need in order to properly respond to your patients’ changing needs. This very effective, yet surprisingly affordable monitoring solution is ideal for pre-operative areas, medical/surgical departments, general floors and beyond.

Note: Mindray accessory kit sold separately, but is necessary for proper function of the unit. If needed, please ensure that the accessory kit option is checked before purchasing.

Features and Benefits

  • 8.4" TFT display, configurable up to 8 waveforms, enables maximum data analysis of your most critical parameters
  • Selectable display modes allow you to adjust data layouts based on particular patient circumstances
  • Comprehensive historical review of all parameters including 96 hours of graphic and list trends, 800 NIBP readings, and 70 alarm episodes
  • 40 seconds of full disclosure waveform recall for detailed review of physiological events
  • 3-trace recorder for printing real-time and historical data is included as standard
  • Compact flash back-up, included on every device, protects your patient data should a sudden loss of power occur
  • Nurse call interface allows connection into most hospital call systems
  • Accessory Kit
    • NIBP Hose, 3.5M
    • Cuff, Reusable-Adult
    • ECG Lead Wire Set - 3 Lead Snap
    • ECG Trunk Cable, 3/5 Lead, Adult/Ped, 6 Pin, Defib Proof
    • 6 Pin Mindray SpO2 Kit (Reusable Adult Finger sensor and Mindray SpO2 cable)
    • Lithium Ion Battery
Standard Features
  • 3/5-lead ECG
  • Non-invasive blood pressure
  • SpO2
  • Pulse rate
  • Respiration
  • Dual-channel temperature
  • Lithium ion battery
  • Integrated recorder
Materials Included
  • CD Operators Manual
  • Line Cord (US)
  • 1 Roll of Paper
  • CF Storage Card
Optional Accessories
  • Wall Mount
  • Rolling Stand
Sample Displays


Technical Specifications

Display
  • Type: 8.4" color TFT
  • Resolution: 800 x 600 pixels
  • Waveforms: 8 selectable
ECG (3 and 5-Lead)
  • Leads: I, II, III, aVR, aVL, aVF, V
  • Gain Selection: x0.125, x0.25, x0.5, x1, x2, auto
  • Sweep Speed: 12.5mm/sec, 25mm/sec, 50mm/sec
  • Bandwidth: Diagnostic Mode: 0.05-100Hz, Monitor Mode: 0.5-40Hz, Surgical Mode: 1-20Hz
  • Defibrillator Overload
  • Protection: Withstand 4000 VAC/50Hz voltage in isolation against electrosurgical interference and defibrillation
  • Recovery Time: <5sec
  • CMRR: Diagnostic Mode: ≥90dB, Monitor Mode: ≥105dB, Surgical Mode: ≥105dB
Heart Rate Meter
  • Measurement Range: Adult: 15-300bpm, Pediatric: 15-350bpm, Neonate: 15-350bpm
  • Accuracy: ±1bpm or ±1%, whichever is greater
  • Resolution: 1bpm
  • Pacer Rejection: When tested in accordance with ANSI/AAMI EC13-1992 Section 4.1.4, the 3/5 Lead ECG derived heart rate meter shall reject all pacer pulses ± 2.0mV to ± 700mV and duration 0.1ms to 2ms with no tail
  • Tall T-Wave Rejection: When tested in accordance with ANSI/AAMI EC13-1992 Section 4.1.2.1 c, the heart rate meter shall reject all T-waves with amplitudes less than 120% of a 1mV, 100ms QRS, and a T wave duration of 180ms and a Q-T interval of 350ms
  • Scaling Signal: 1mV ±5%
Respiration
  • Range: Adult: 0-120bpm, Pediatric/Neonate: 0-150bpm
  • Resolution: 1bpm
  • Accuracy: 7-150bpm: ±2bpm or ±2%, whichever is greater
  • 0-6bpm: undefined
  • Lead: I or II (default: lead II)
  • Sweep Speed: 6.25mm/sec, 12.5mm/sec, 25mm/sec
Pace Pulse
  • Pulse Indicator: Pace pulses meeting the following conditions are marked by the PACE indicator
    • Amplitude: ±4 to ±700mV (3/5-lead)
    • Width: 0.1 to 2ms
    • Rise time: 10 to 100μs
  • Pulse Rejection: When tested in accordance with the ANSI/AAMI EC13-2002: Sections 4.1.4.1 and 4.1.4.3, the heart rate meter rejects all pulses meeting the following conditions
    • Amplitude: ±2 to ±700mV
    • Width: 0.1 to 2ms
    • Rise time: 10 to 100μs
    • Minimum input slew rate: 20V/s RTI
Non-Invasive Blood Pressure
  • Measurement Method: Oscillometric
  • Measurement Modes: Manual, interval, continuous
  • Connector Type: Rectus
  • Units of Measure: mmHg, kPa (user-selectable)
  • Resolution: 1mmHg
  • Systolic Range: Adult: 40-270mmHg, Pediatric: 40-200mmHg, Neonate: 40-135mmHg
  • Diastolic Range: Adult: 10-210mmHg, Pediatric: 10-150mmHg, Neonate: 10-100mmHg
  • Mean Range: Adult: 20-230mmHg, Pediatric: 20-165mmHg, Neonate: 20-110mmHg
  • Accuracy: Mean error: < ±5mmHg, Standard deviation: < 8mmHg
  • Cuff Deflation Technique: Step bleed
  • Cuff Inflation: Volume of 500cc to 300mmHg in <20sec
  • Over Pressure Protection: Double safety protection (hardware and software)
  • Pulse Rate Range: 40-240bpm
  • Pulse Rate Accuracy: ±3bpm or ±3%, whichever is greater
Invasive Blood Pressure
  • Measurement Range: -50 to 300mmHg
  • Resolution: 1mmHg
  • Accuracy: 1mmHg or ±2%, whichever is greater
  • Zero Offset Range: ±200mmHg
  • Excitation: 5VDC, ±2% Minimum load resistance will be 300Ω per transducer
  • Frequency Response: DC to 12.5Hz ±1Hz, -3db
  • Waveform Scales:
    • ART: 0 to 300mmHg
    • PA: -6 to 120mmHg
    • CVP: -10 to 40mmHg
    • RAP: -10 to 40mmHg
    • LAP: -10 to 40mmHg
    • ICP: -10 to 40mmHg
    • IBP1/IBP2: -50 to 300mmHg
Pulse Rate from Invasive Blood Pressure
  • Measurement Range: 25-350bpm
  • Resolution: 1bpm
  • Accuracy: 25-200bpm: ±1bpm or ±1%, whichever is greater.
  • 201-350bpm: ±2%
Pulse Oximetry
  • Measurement Range: 0-100%
  • Resolution: 1%
  • Accuracy: ±2% (70-100%, Adult/Pediatric, no motion)
  • ±3% (70-100%, Neonate, no motion)
  • ±3% (70-100%, Adult/Pediatric/Neonate, motion)
  • 0-69% unspecified
Pulse Rate with Mindray SpO2
  • Measurement Range: 20-254bpm
  • Resolution: 1bpm
  • Accuracy: ±3bpm (no motion), ±5bpm (motion)
Temperature
  • Scale: Selectable °C or °F
  • Channels: 2
  • Measurement Range: 0°C to 50°C (32°F to 122°F)
  • Resolution: 0.1°C
  • Accuracy: ±0.1°C (excluding sensor) ±0.2°C (including YSI 400 series sensor)
  • Measurement Time:
    • Body surface: <100sec (using YSI 400 series sensor)
    • Body cavity: <80sec (using YSI 400 series sensor)
Data Storage
  • Trend Data: 96 hours at selectable resolutions: 1min, 5min, 10min - 1 hour at selectable resolutions: 1sec, 5sec
  • Alarm Events: 70 alarm events and associated waveforms (selectable waveform lengths: 8sec, 16sec or 32sec)
  • NIBP Measurements: 800 (systolic, diastolic, mean pressure and measurement time)
Recorder
  • Type: Thermal array
  • Speed: 25mm/sec
Battery
  • Type: Rechargeable lithium ion
  • Run Time: 3 hours using a new, fully charged battery and monitoring ECG, SpO2 and NIBP measurements every 15min at 25°C
  • Recharge Time: 6.5 hours
Interfacing
  • Connectors: 1 AC power connector, 1 RJ45 network connector, 100 BASE-TX, 1 VGA connector, 15-PIN D-sub, 1 BNC connector, 1 equipotential grounding connector
Physical Dimensions
  • Monitor Size: 24cm(H) x 26.1cm(W) x 17.1cm(D) 9.4"(H) x 10.3"(W) x 6.7"(D)
  • Monitor Weight: Less than 5kg (11lbs) standard configuration (ECG, Resp, SpO2, 2-Temperature, NIBP) with 1 lithium ion battery
Environmental
  • Operating Temperature: 0°C to 40°C
  • Storage Temperature: -20°C to 60°C
  • Operating Humidity: 15% to 95%, non-condensing
  • Storage Humidity: 10% to 95%, non-condensing
  • Operating Altitude: -500 to 4600m (-1640 to 15092ft)
Power Requirements
  • AC Voltage: 100–240VAC, 50/60Hz
  • Power: 110VA
Safety
  • Type of Protection: Class I with internal electric power supply
  • Degree of Protection: ECG/Resp/Temp/SpO2/NIBP/IBP: CF
  • Protection Against Ingress of Fluids: Not protected
Documentation

Share your knowledge of this product with other customers... Be the first to write a review

Browse for more products in the same category as this item:

Vital Sign Monitors
Vital Sign Monitors > Mindray Monitors