This product has been discontinued. For a comparative product, click here for the Propaq® LT Monitors.
Advanced surgical techniques, new anesthetics, the need to lower
healthcare costs, they're all pointing toward more outpatient
surgeries and procedures in hospitals, surgical centers and physician
offices. That's why you need the Welch Allyn Atlas monitor, a vital
signs monitor specifically designed to monitor patients during conscious
sedation and IV anesthesia procedures for these markets.
Designed
with input from physicians and nurse anesthetists, this conscious
sedation monitor offers bright, clear,
across-the-room
visibility with large LED readouts and crisp waveform displays, giving
you at-a-glance convenience and more time to concentrate on your
patient. The intuitive design means users can learn to operate the
monitor in just minutes.
The Atlas monitor goes where your
patient goes, effortlessly. Lightweight and portable, it can be set on a
cart or countertop, wheeled on its own dedicated mobile stand, mounted
on the wall or carried using the integrated side handle.
We offer
the flexibility to configure the monitor that best fits your needs, and
it's upgradeable, should your needs change in the future.
Designed specifically for conscious
sedation and IV anesthesia in a wide variety of healthcare sites.
Features and Benefits
- Hard-working, easy-to-use monitor with configuration options to meet your individual needs.
-
All monitors include 3- or 5-lead ECG, Nellcori pulse oximetry (SpO2) and NIBP.
-
Optional parameters include temperature, respiration, EtCO2 and integrated printer
(622 or 623 series)
-
Bright, crisp waveforms and multicolored LED displays for across-the-room visibility.
-
Simple, straightforward menu operation.
Technical Specifications3-Lead or 5-Lead Selectable ECG
- Heart Rate Range: 20-250 bpm.
- Heart Rate Accuracy: ±3 bpm or ±3%, whichever is greater.
- Bandwidth: 0.5-40 Hz (monitor mode) or 0.05-100 Hz (extended mode).
- Display Sweep Speeds: 25, 12.5 or 6.25 mm/sec.
Noninvasive Blood Pressure
- Systolic Measurement Range: 60-250 mmHg.
- Diastolic Measurement Range: 30-160 mmHg.
- Blood Pressure Accuracy: Meets AAMI SP10-1992 Standard.
- Operating Modes: Manual (single determination) or Automatic Mode.
- Cycle Interval Choices: 1, 3, 5, 10, 15, 30 or 60 minutes.
- Pulse Oximetry (Nellcor Puritan Bennett™)
- Measurement Range: 40-100%
- Accuracy: ±2 digits (no motion), ±3 digits (with motion) in the range of 0-100%
- Pulse Rate Range: 20-250 bpm.
- Pulse Rate Accuracy: ±3 digits (no motion), ±5 digits (with motion)
Temperature
- Range: 17-50° C (62.6-122° F)
- Accuracy: Meets ASTM Accuracy Standards.
- Probe Type: YSI Series 400 Compatible.
Impedance Respiration
- Range: 5-100 breaths per minute.
- Accuracy: ±3 breaths per minute.
End Tidal Carbon Dioxide
- Measurement Technique: Sidestream.
- Measurement Range: 0-98 mmHg.
- ETCO2 Accuracy: 0-40 mmHg ±3 mmHg; 41-98 mmHg ±10% of reading.
- Breath Rate Measurement Range: 5-100 breaths per minute.
- Breath Rate Accuracy: ±1 breath per minute or ±5%, whichever is greater.
General
- Size: 35 cm x 23 cm x 25 cm (13.8” x 9.1” x 9.8”)
- Weight: 4.3-5.5 kg (9.5-12.1 lbs) dependent on configuration.
- Power Requirements: 120 VAC, 50/60 Hz; 240 VAC, 50/60 Hz.
- Printer: Thermal array, prints waveforms and numerical trends.
- Battery: Lead acid, rechargeable.
Warranty Information