HVL Filters and Holder
HVL Filters and Holder


 
Discontinued
This product has been discontinued. Click here to return to Mammography products.

HVL filters help the medical physicist or service engineer confirm the necessary filtration required to remove the damaging low-energy radiation from the primary beam of the various X-ray tubes used in diagnostic radiology. The L435 HVL Holder is designed to simplify the routine HVL measurement process. It eliminates the need to use tape to attach the HVL filters to the collimator housing.

Permanently bonded to the center of the base plate is an acrylic pocket, open on one side and designed to hold a standard or high purity Al filter set. The polycarbon material is easily cut with a standard utility knife or sheers to accommodate the two most common collimator track sizes in a given department. The base may also be attached with hook-andloop type fastener strips for odd sized collimators. In either case, the filters themselves are protected from damage associated with the application and removal of heavy medical/ surgical tape.

Note: This product cannot be exported to the following countries: United Kingdom, France, New Zealand, Germany, Austria, Italy, Japan, and Australia.

Features and Benefits

Model L430 - Standard Aluminum Set
  • Standard high-purity aluminum for HVL determinations of mid-range generators (80-140 kVp). 
  • Includes 11 filters (10 cm x 10 cm); five of them with 1.0 mm thickness, two of them with 0.5 mm thickness and four of them with 0.1 mm thickness. 
  • Made with type-1100 aluminum. 
  • Weight: 0.06 kg (15 lb)
Model L431 - Copper HVL Filter Set
  • For HVL determinations of high-range X-ray generators (140-400 kVp). 
  • Includes 10 filters (10 cm x 10 cm): four of them with 1.0 mm thickness, two of them with 0.5 mm thickness, and four of them with 0.1 mm thickness.
  • Weight: 0.56 kg (1.25 lb)
Model L434 - Standard Aluminum Set
  • The ultra-high purity aluminum typically required for HVL determination of mammography. 
  • Because type 1100 aluminum is only 99.0% pure, it has some impurities that can give an HVL value that is 7.5% lower than those measured with pure aluminum. 
  • When doing HVL measurements on a mammography unit, it is recommended that the highest purity aluminum be used. This set of six attenuators satisfies this recommendation, because they are 99.9% pure (type 1145 aluminum).
  • Weight: 0.56 kg (1.25 lb)
Model L435 - HVL Filter Holder
  • Base Dimensions: 24.1 cm x 24.1 cm  (9.5" x 9.5")
  • Folder Pocket Dimensions: 10.2 cm x 10.2 cm x 9 mm  ("4 x 4" x 0.35")
  • Weight: 0.15 kg (0.3 lb) 



Share your knowledge of this product with other customers... Be the first to write a review

Browse for more products in the same category as this item:

Radiology > Tissue Simulation & Phantom
Radiology > Tissue Simulation & Phantom > Mammography
 
Welch Allyn Authorized Distributor Volusion Secure Site ABOUT TRUST ONLINE Medical Device Depot is a ShopWiki Approved Store Become.com DOTmed Certified: Medical Device Depot PayPal Verified
BBB Accredited Business Physio Control Payment Options Compare prices at WillyFogg.com