The Advisor Vital Signs Monitor offers flexibility in multi-parameter monitoring. Core parameters and a host of options allow the monitor to be tailored to patients' needs, neonate to adult.

Combine the intuitive interface, integrated carrying handle and high-resolution display with the optional rechargeable battery and dual trace printer to create a highly versatile monitoring solution.

Features and Benefits
  • Basic parameters include 3-lead ECG, respiration, NIBP and SpO2.
  • Optional parameters of 5-lead ECG, invasive pressure/temperature and capnography.
  • 10-inch color display.
  • Large, bright color display.
  • Multiple options.
  • NOW COMPATIBLE with ProVation Medicals MultiCaregiver information management system which allows data collected by the Advisor during procedures such as endoscopies and minor surgeries to be downloaded in real time to MultiCaregivers procedure log and subsequently interfaced with a facilitys electronic medical records.


Item# Item Name List Price Our Price Qty Add
9200-3 Advisor 2 Hour Battery without CO2
$250.00
$200.00
(Out of Stock)
9200-5 Advisor Nellcor Oximetry (includes #DEC8 Extension Cable & #DS100A Adult Reusable Sensor)
$600.00
$450.00
(Out of Stock)
9200-2 Advisor Recorder
$750.00
$600.00
(Out of Stock)
9200 Advisor Vital Signs Monitor Base Unit
$4,850.00
$4,000.00
(Out of Stock)
9200-1 Advisor Vital Signs Monitor Base Unit
$300.00
$300.00
(Out of Stock)
Check the items you wish to purchase, then click


Technical Specifications:
  • Displays & Controls
    Main Display
    10.4"
  • high-resolution active matrix color LCD
    Resolution = 640 x 480
  • pixels
    Function Keys
    ON/OFF, BP Zero All, NIBP start/stop, PRINT
    start/stop,
  • Alarm Silence, rotary encoded
    push-button control knob
    Indicators
  • AC Power
    Green LED w/power plug icon
    Battery Power
    Green LED
  • w/battery icon
    SPO2
    Range
    0-100% (functional saturation)
  • Accuracy
    2% @ 70-100% < 70% unspecified
    Alarm Range
    50-100%
  • and OFF
    Averaging
    4, 8, 16 pulse beat average
    Pulse Rate
  • (SpO2) or Heart Rate (ECG)
    Range
    30-250 bpm (SpO2) 30-300 bpm
  • (ECG)
    30-180 bpm (NIBP) 30-300 bpm (IBP-ART)
    Accuracy
    2 bpm
  • or 2% (whichever is greater)
    Alarm Range
    30250 bpm and OFF
  • Averaging
    Fixed at 8 seconds
    ECG
    Configuration
    3-lead
  • (standard) or 5-lead (optional)
    Lead Selection
    I, II, III, aVL,
  • aVR, aVF and V
    Display Gain
    x1/2, x1, x2, x4
    Defibrillation
  • Protection
    Yes
    Pace Detect/Reject
    Yes
    Sweep Speed
    25
  • mm/sec
    Impedance Respiration
    (not available in neonate mode)
  • Measuring Method
    Impedance Pneumography
    Rate
    3-150
  • breaths/min
    Accuracy
    1 breath/min or 5% (whichever is greater)
  • Alarm Range
    3-120 bpm and OFF (1 bpm steps)
    Non -Invasive Blood
  • Pres ure
    Range
    Adult/Pediatric: 20-250 mmHg
    Neonate: 20-130
  • mmHg
    Accuracy
    Meets or exceeds SP10-1992 Standard for
    NIBP
  • accuracy
    Cuff Pressure Accuracy
    3 mmHg or 2% (whichever is
  • greater)
    Inflation Pressure Adjustment Range
    Adult/Pediatric:
  • 50-250 mmHg
    Neonate: 50-120 mmHg
    Default Inflation
    Adult:175
  • mmHg
    Pediatric:150 mmHg
    Neonate: 100 mmHg
    Pressure Alarm
  • Range
    Adult/Pediatric: 20-250 mmHg and OFF
    (1 mmHg steps)
    Neonate:
  • 20-135 mmHg and OFF
    (1 mmHg steps)
    2 Channe l Invasive Pressure
  • optional
    Range
    -50 mmHg 300 mmHg (after zeroing)
  • Accuracy
    1 mmHg or 1% (whichever is greater)
    (exclusive of
  • transducer error)
    Alarm Range
    -50 300 mmHg (in 1 mmHg steps)
    2
  • Chane l Temperature optional
    Range
    32F-122F (0C-50C)
  • Accuracy
    0.2F (0.1C) plus temp sensor tolerance
    Resolution
    0.1F
  • (0.1C)
    Alarm Range
    68F-122F (20.0C-50.0C) and OFF
  • Sensor Compatibility
    YSI 400 or equivalent
    Sidestream Capnography
  • optional
    (not available in neonate mode)
    Measurement
    Non-Dispersive
  • IR absorption
    CO2 Measurement Range
    0-10% CO2 STPD
    Display
  • Range
    0-100 mmHg; 0-13.3 kPa; 0-10% CO2
    Accuracy
    2 mmHg or
  • 4% of reading
    (whichever is greater) (0-10% CO2)
    Alarm Range
    0-100
  • mmHg; 0-13.3 kPa; 0-10% CO2 and OFF
    N2O Compensation
    Selectable
  • 40% (default = OFF)



Share your knowledge of this product with other customers... Be the first to write a review