Home > Tissue Simulation & Phantom >

Pro-Project Pro-NM MultiAlign
Pro-Project Pro-NM MultiAlign


 
Additional Views:


List Price: $4,446.00
Our Price: $3,705.00
You save $741.00!




Product Code: 08-701

Details
 
The Phantom designed for a simple and cost effective verification of image alignment and distortion in hybrid scanning systems like PET/CT or NM/CT. It consists of a cylinder that can be filled with a variety of fluids. Several non-parallel rods of varying diameter and at certain angles in relation to the phantom’s axes run the entire length of the cylinder. Images produced with different modalities allow quick and simple identification of misalignments on fused studies.

Additionally, the phantom contains 2 cm calibration module that is used by Diagnomatic software to accurately identify the phantom’s position in space and accurately calculate any mismatches.

Features and Benefits
  • Complies with:
    • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
  • CE certified
  • The manual provides detailed guidelines for carrying out each test, results assessment and registration
  • Filling plugs and vent ports
  • Carrying handle
Technical Specifications
  • PMMA Cylinder Dimensions: 250 x 260 x 410 mm
  • 20 mm thick calibration section with four through holes
  • 4 PMMA Rods: 30, 25, 20, 15 mm in diameter
Warranty Information
  • 1-Year Warranty

Share your knowledge of this product with other customers... Be the first to write a review

Browse for more products in the same category as this item:

Tissue Simulation & Phantom