Contrast details mammography phantom according to European Protocol for quality assurance in Digital Mammography. It may be used to determine if mammographic images indicate objects with very low contrast and very small diameter, to find out the optimum exposure technique, or to compare image quality at various object thicknesses by variation of the amount of PMMA thickness.
Features and Benefits
- Unmatched accuracy – up to 30 times better than other products on the market
- Gold disc objects positioned as per EUREF specs
- Objects diameter, thickness and layout can be modified according to individual specifications
- No materials other than PMMA are used, so there is no need to calculate PMMA-equivalent attenuation for other materials such as aluminium
- For full-field analog and digital units
- Main module contains gold (99.99%) discs organized in a 16 x 16 matrix (diameter x thickness) placed at a depth of 10 mm
- One or multiple discs per cell configuration
- Discs have following standard diameters: 0.05, 0.1, 0.15, 0.2, 0.25, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9, 1.0, 1.25, 1.5 and 2.0 mm
- Discs have following standard thicknesses: 30, 40, 50, 60, 70, 90, 120, 150, 200, 250, 350, 500, 700, 1000, 1700 and 2000 mm
- 2 additional 240 x 180 x 10 mm PMMA plates
- 2 additional 240 x 180 x 5 mm PMMA plates
- Carrying case is included
Technical Specifications - Dimensions: 9" x 7" x 0.78"
- Diameter Accuracy: 0.001 mm (1µm)
- Thickness Accuracy: 0.1 mm (0.0001 µm)
Warranty Information
|
|
|
|
|
|
|