Home > Pulse Oximeters/Capnographs > Hand-Held Oximeters >

MiniCorr Digital Hand-Held Pulse Oximeter System
MiniCorr Digital Hand-Held Pulse Oximeter System


 




Product Code: 3402
Discontinued

Details
 
This product has been discontinued. For a comparable product, click here for the Nellcor™ Portable SpO2 Patient Monitoring System w/ Adult, Pediatric & Neonatal

The BCI MiniCorr 3402 pulse oximeter puts pulse oximetry technology in the palm of your hand.

Designed to provide fast, reliable SpO2, Pulse Rate and Pulse Strength measurements, the MiniCorr is ideally suited for spot-checking or bedside monitoring in home care and clinical environments, as well as during sleep screenings.

Features and Benefits
  • Reliably measure SpO2, Pulse Rate and Pulse Strength.

  • SAC Technology informs operator of excess artifacts and improper sensor positioning.

  • Easy-to-use, 5 key operation.

  • Infrared link to printer.

  • Lightweight for easy transport.

  • Battery power or battery bypass.

Materials Included
  • Adult Finger Sensor

  • Manual

  • 6 AA batteries

Technical Specifications

Displays, Indicators & Keys
  • SpO2 LED numeric display: 0.3 inch (7.62mm) high
  • Pulse Rate LED numeric display: 0.3 inch (7.62mm) high
  • Pulse Strength: Logarithmically scaled 8-segment LED bar graph
  • Low Battery LED: lights when approximately 30 minutes of battery life remains
  • Audio Disabled LED: lights to indicate temporary or indefinite status
  • Sensor LED: lights to indicate an Oximeter sensor problem
  • Artifact LED: lights to indicate the presence of excess motion or other artifact
  • Power Green LED: indicates monitor is connected to AC power
  • Keys: On/Off; Alarm Select, Alarm Silence; UP arrow - Alarm Limit/Pulse Volume Adjustment; DOWN arrow - Alarm Limit/ Pulse Volume Adjustment; Print/Clear

SpO2:

  • Range: 0-99% functional SpO2 (1% increments)
  • Accuracy: ±2% at 70-99%; ±3% at 50-69%
  • Averaging: 4, 8, or 16 pulse beat average
  • Alarm Range:
    • > High - 50%-99% & OFF (1% steps)
    • > Low - OFF & 50%-99% (1% steps)

Pulse Rate:

  • Range: 30-254 bpm
  • Accuracy: ±2% or 2bpm, whichever greater at 30-254 bpm
  • Alarm Range:
    • > High - 5-250 bpm & OFF (5 bpm steps)
    • > Low - OFF & 5-250 bpm (5 bpm steps)
  • Averaging: 8 or 16 second average

Trend Storage

  • Parameters: SpO2, Pulse Rate, Pulse Strength, Relative Time
  • Interval: 4 seconds to 30 seconds
  • Capacity: 12 hours of storage at 4 sec. interval
  • 24 hours of storage at 8 sec. interval
  • 90 hours of storage at 30 sec. interval

Power

  • Battery 6 standard AA alkaline cells
  • Battery Life 24 hours without printing
  • AC Power (optional) External 12 VDC power supply

Printer Output:

  • Data Log Mode: SpO2, and Pulse Rate printed every 5 seconds.
  • Trend Data Mode: Up to 90 hours of previously stored data, collected from 1 to 99 patients.
  • PC Communications: RS232C With the oximeter sensor removed, trend data can be transmitted from the sensor connector to PC serial port by way of a null modem cable.

Physical:

  • Dimensions 6.6” H x 2.75” W x 1.43” D, 16.7 cm x 7.0 cm x 3.6 cm
  • Weight: 16 oz. (454 grams) with batteries

Environment:

  • Operating Temperature: 32°F to 131°F (0°C to 55°C)
  • Storage Temperature: -40°F to 167°F (-40°C to 75°C)
  • Relative (non-condensing): Humidity 15-95% (storage), 10-95% (operating)

Standards:

  • EN60601-1 (IEC 60601-1), UL544, CSA Std C22.2
  • No. 125-1984, EN60601-1-1, EMC: EN60601-1-2