Features and Benefits
- 12.1’’ LCD display, high resolution,
- Large number display for long distance vision
- Digital SPO2 technology offers anti-motion and anti-low perfusion funtions
- Up to 66 beds network capacity to connect with HIS
- 168 hours tabular and graphic trends & data storage
- Drug dose calculation
- OXYCRG software
- Bed to bed view
Technical Specifications
General
- Safety: IEC60601 (GB9706)
- Dimension: 318mm (W) x 264mm (H) x 152mm(D)
- Power Supply : 100~240V; AC: 50~ 60 Hz
- Alarm Indication: Visual and audible alarm
- 1800 groups of alarm events review
- Application: Adult, Pediatric and neonate
ECG- Leads: I, II, III, aVR, aVL, aVF, V
- Gains: 2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 40mm/mV, auto
- Input impedance: =5MO (d.c.~60Hz)
- Common Mode Rejection Ratio: =85dB(diagnostic mode)
- =105dB(monitoring & surgica mode)
- Bandwith: 0.5-40Hz (monitoring mode)
- 0.05-120Hz(diagnostic mode)
- 1-25Hz (surgical mode)
- Protection: Segregation breakdown voltage 4000VAC, 50/60Hz
- Resistant to high-frequency electrotome, myoelectricity, AC, drifting interference
- Indication of Detachment of electrodes: Available
- Playback of Cardiogram: Available
- Scanning Speed: 12.5mm/s, 25mm/s, 50mm/s
- Heart Rate Range:10-300BPM
- Accuracy: ±1BPM
- ECG waveform playback: 30 minutes
NIBP
- Method: Automatic oscillometric
- Operation Modes: Singles (manual), Cycle (automatic), Rapid (continuous)
- Measurement : Systolic pressure/diastolic pressure/mean pressure
- Range: Adult: 10-270mmHg
- Pediatric: 10-235mmHg
- Neonate: 10-135mmHg
- Unit: mmHg / Kpa selectable
- Auto Measurement: 1,2,3,4,5,10,15,20,30,45,60,90 minutes, 2,4,8, hours
- 1000 groups of NIBP measuring data review
RESP - Method: Impedance Measurement
- Measuring Range: 0-120bpm
- Alarm Limit: 0-120bpm, upper and lower limit continuously adjustable
- Apnea alarm: Available
BLT Digital SPO2 (Pulse Oximetry)
- Range: 0-100%
- Accuracy: ±2% (70%~100%);
- ±3%(35%~ 69%)
- Unspecified (0~34%)
- Range & Accuracy of Pulse Rate : 20~250BPM ± 1BPM
- Channel : 2 channels
- Range : 0~50°
- Unit : Celsius, Fahrenheit
- Resolution : 0.1°
- Accuracy : ±0.1° (25°~50°)
- Alarm Range : 0~50°
IBP
- Channel : 2 channels
- Range : -50~300 mmHg
- Resolution : 1 mmHg
- Accuracy : ±2 mmHg (±2%) without probe, ±4 mmHg (±2%) with probe
- Unit : mmHg , Kpa
- Converter sensitivity: 5 uV/V/mmHg
- Names of pressures: ART/PA/CVP/LAP/RAP/ICP
EtCO2 (side stream)
- Method: Infrared spectrum
- Range: 0.0-13.1%
- Resolution : 1 mmHg (0.1%)
- Accuracy : <5% EtCO2, ±0.3%
- >5% EtCO2, <10% (of Readings)
EtCO2 (main stream)
- Method: Infrared spectrum
- Range: 0.0-10% (0-76mmHg)
- Resolution : 1 mmHg (0.1%)
- Accuracy : 5% (±4.0 mmHg) or <10% (of Readings)
IRMA multi-gas (main stream)- Measure method: Infrared spectrum
- Fi and ET values: CO2,N2O,O2,agent(HAL, ISO, ENF, SEV, DES)
- Resolution: 1mmHg
- Unit: %, mmHg
- Calibration: Room air calibration performed automatically when changing airway adapter (<5 sec)
- Warm-up time: 5s
- Measuring and alarm range of multi-gas:
- CO2: 0-10 % ,±0.3% or ±6% of reading, whichever is greater
- N2O: 0-100 %, ±2% or ±8% of reading, whichever is greater
- O2: 0-100 %, ±3 %
- HAL, ISO, ENF: 0-5%, ±0.15% or ±8% of reading, whichever is greater
- SEV: 0-8%, ±0.15% or ±8% of reading, whichever is greater
- DES: 0-18%, ±0.15% or ±8% of reading, whichever is greater
- Respiration rate Measuring alarm range and accuracy: 3~100 bpm, ±1 bpm
- Alarm indication: Sound and light indication
Environment Conditions- Temperature:
Operation 0°-40°
- Storage / Transportation : -20°-50°
- Humidity:
Operation: =85% noncondensing
- Storage / Transportation : =93% noncondensing
- Atmospheric Pressure:
Operation: 860~1060hPa
- Storage / Transportation : 500~1060hPa
Technical Specifications
|
|
|
|
|
|
|