DELTA 3000A - Defibrillator and Transcutaneous Pacer Analyzer
DELTA 3000A - Defibrillator and Transcutaneous Pacer Analyzer


 


Product Code: 650-A
Discontinued

Details
 
This product has been discontinued. For a comparable product, click here for the Delta 3300 - Defibrillator/Transcutaneous Pacemaker Analyzer


The Delta 3000A is a compact instrument that tests defibrillators and transcutaneous pacemakers. It is also a 12 lead ECG arrhythmia simulator.

The Delta 3000A tests defibrillators with monophasic or bi-phasic waveforms and all AEDs. It will test transcutaneous pacemkaers and includes 12 built-in user selectable test loads. The 12 lead ECG simulator includes 12 arrhythmias plus sine, square, and triangle performance waveforms.

The Delta 3000A may be easily upgraded to add the capability of testing transvenous pacemakers.

Features and Benefits
  • Compact, Lightweight Case
  • Simple, Menu Driven Operation
  • Large Easy to read LCD Display
  • Simultaneous Display of All Test Measurements
  • Dedicated Print Key for Immediate Documentaion
  • RS 232 Interface
  • Battery or AC Operation
Measurements Tested
  • Measures: Energy, Peak Voltage, Peak Current, and Cardioversion
  • Storage and Playback of output waveforms
  • Built-in 50 Ohm Test Load
  • Built-in User Selectable Test Loads to 1000 Ohms
  • Built-in 50 Ohm Test Load
  • Amplitude, Rate, Pulse Width Accuracy Tests
  • Sensed and Paced Refactory Period Tests
  • 50 and 60 Hz Immunity Tests
  • Measures Pulse Energy
Materials Included
  • Delta 3000 Defibrillator/Pacer Analyzer
  • Hard carrying case
  • Paddle adapters
  • Two sets of universal connectors
  • AC adapter
  • Batteries
  • Instruction manual
  • Certificate of calibration
Technical Specifications

Defibrillator Analyzer
  • Energy Measurement
    • Load: 50 Ohm +/- 1 %
    • Range (High): 900 J, Res.1 J
    • Resolution: 1 Joules
    • Accuracy: +/- 2% of the reading + 1 J for Hi Range or + 0.5 J for Low Range
    • Maximum Voltage: 5200 V
    • Maximum Current: 100 Amp (H)
    • Sync. Time: 0 - 250 ms
    • Waveform Output - Real Time:
      • Output 1000: 1 Stored.
      • Discharge waveform expanded 200 : 1 and is repeated 8 sec. It can be viewed through the paddles and through the ECG High Level Output.
Pacemaker Analyzer
  • Transcutaneous:
    • Test Load: 50 Ohms fixed; variable 100 to 1000 Ohms user selectable
    • Current: 1 to 200 mA
    • Accuracy: +/- 5% + 0.5 mA
    • Pulse Width: 0.5 to 80 ms
    • Accuracy: +/- 1% + 0.1 ms
    • Refractory: 20 - 500 ms
    • Accuracy: +/- 2% or + 10 ms
ECG Arrhythmia Simulator
  • ECG: Normal Sinus Rhythm. Rate: 30, 60, 90, 120, 240
  • Performance: Sine, Square, Triangle and Pulse Rate: 0.5, 1, 2, 10, 15, 20, 25, 40Hz
  • Arrhythmias: VFBC, VFBF, VTAC, AFIB, PVC, PVC1, PVC2, BGY, RBB, AFIB, ATRFT, NSR
  • Waveform Output: 12 Lead ECG and Paddle and High Level Output Jacks
Warranty Information
  • 1-Year Warranty