Home > Vital Sign Monitors >

Advisor Vital Signs Monitor Base Unit
Advisor Vital Signs Monitor Base Unit


 
This product has been discontinued. Click here to return to the Vital Signs Monitors.
The Advisor® Vital Signs Monitor offers flexibility in multi-parameter monitoring. Core parameters and a host of options allow the monitor to be tailored to patients' needs, neonate to adult.

Combine the intuitive interface, integrated carrying handle and high-resolution display with the optional rechargeable battery and dual trace printer to create a highly versatile monitoring solution.

Features and Benefits
  • Basic parameters include 3-lead ECG, respiration, NIBP and SpO2.
  • Optional parameters of 5-lead ECG, invasive pressure/temperature and capnography.
  • 10-inch color display.
  • Large, bright color display.
  • Multiple options.
  • NOW COMPATIBLE with ProVation® Medical’s MultiCaregiver information management system which allows data collected by the Advisor® during procedures such as endoscopies and minor surgeries to be downloaded in real time to MultiCaregiver’s procedure log and subsequently interfaced with a facility’s electronic medical records.


Item# Item Name List Price Our Price Qty Add
9200-3 Advisor 2 Hour Battery without CO2
$250.00
$200.00
(Out of Stock)
9200-5 Advisor Nellcor Oximetry (includes #DEC8 Extension Cable & #DS100A Adult Reusable Sensor)
$600.00
$450.00
(Out of Stock)
9200-2 Advisor Recorder
$750.00
$600.00
(Out of Stock)
9200 Advisor Vital Signs Monitor Base Unit
$4,850.00
$4,000.00
(Out of Stock)
9200-1 Advisor Vital Signs Monitor Base Unit
$300.00
$300.00
(Out of Stock)
Check the items you wish to purchase, then click


Technical Specifications:
  • Displays & Controls
    • Main Display
    10.4"
  • high-resolution active matrix color LCD
    Resolution = 640 x 480
  • pixels
    • Function Keys
    ON/OFF, BP Zero All, NIBP start/stop, PRINT
    start/stop,
  • Alarm Silence, rotary encoded
    push-button control knob
    Indicators
  • AC Power
    Green LED w/power plug icon
    • Battery Power
    Green LED
  • w/battery icon
    SPO2
    • Range
    0-100% (functional saturation)
  • Accuracy
    ±2% @ 70-100% < 70% unspecified
    • Alarm Range
    50-100%
  • and OFF
    • Averaging
    4, 8, 16 pulse beat average
    Pulse Rate
  • (SpO2) or Heart Rate (ECG)
    • Range
    30-250 bpm (SpO2) 30-300 bpm
  • (ECG)
    30-180 bpm (NIBP) 30-300 bpm (IBP-ART)
    • Accuracy
    ±2 bpm
  • or ±2% (whichever is greater)
    • Alarm Range
    30–250 bpm and OFF
  • Averaging
    Fixed at 8 seconds
    ECG
    • Configuration
    3-lead
  • (standard) or 5-lead (optional)
    • Lead Selection
    I, II, III, aVL,
  • aVR, aVF and V
    • Display Gain
    x1/2, x1, x2, x4
    • Defibrillation
  • Protection
    Yes
    • Pace Detect/Reject
    Yes
    • Sweep Speed
    25
  • mm/sec
    Impedance Respiration
    (not available in neonate mode)
  • Measuring Method
    Impedance Pneumography
    • Rate
    3-150
  • breaths/min
    • Accuracy
    ±1 breath/min or ±5% (whichever is greater)
  • Alarm Range
    3-120 bpm and OFF (1 bpm steps)
    Non -Invasive Blood
  • Pres ure
    • Range
    Adult/Pediatric: 20-250 mmHg
    Neonate: 20-130
  • mmHg
    • Accuracy
    Meets or exceeds SP10-1992 Standard for
    NIBP
  • accuracy
    • Cuff Pressure Accuracy
    ±3 mmHg or ±2% (whichever is
  • greater)
    • Inflation Pressure Adjustment Range
    Adult/Pediatric:
  • 50-250 mmHg
    Neonate: 50-120 mmHg
    • Default Inflation
    Adult:175
  • mmHg
    Pediatric:150 mmHg
    Neonate: 100 mmHg
    • Pressure Alarm
  • Range
    Adult/Pediatric: 20-250 mmHg and OFF
    (1 mmHg steps)
    Neonate:
  • 20-135 mmHg and OFF
    (1 mmHg steps)
    2 Channe l Invasive Pressure –
  • optional
    • Range
    -50 mmHg – 300 mmHg (after zeroing)
  • Accuracy
    ±1 mmHg or ±1% (whichever is greater)
    (exclusive of
  • transducer error)
    • Alarm Range
    -50 – 300 mmHg (in 1 mmHg steps)
    2
  • Chane l Temperature – optional
    • Range
    32°F-122°F (0°C-50°C)
  • Accuracy
    ±0.2°F (±0.1°C) plus temp sensor tolerance
    • Resolution
    0.1°F
  • (0.1°C)
    • Alarm Range
    68°F-122°F (20.0°C-50.0°C) and OFF
  • Sensor Compatibility
    YSI 400 or equivalent
    Sidestream Capnography –
  • optional
    (not available in neonate mode)
    • Measurement
    Non-Dispersive
  • IR absorption
    • CO2 Measurement Range
    0-10% CO2 STPD
    • Display
  • Range
    0-100 mmHg; 0-13.3 kPa; 0-10% CO2
    • Accuracy
    ±2 mmHg or
  • ±4% of reading
    (whichever is greater) (0-10% CO2)
    • Alarm Range
    0-100
  • mmHg; 0-13.3 kPa; 0-10% CO2 and OFF
    • N2O Compensation
    Selectable
  • 40% (default = OFF)



Share your knowledge of this product with other customers... Be the first to write a review

Browse for more products in the same category as this item:

Vital Sign Monitors
Vital Sign Monitors > Smith BCI Monitors