BCI MiniCorr 3402 pulse oximeter puts pulse oximetry
technology in the palm of your hand.
Designed to provide
fast,
reliable SpO2, Pulse Rate and Pulse Strength measurements,
the MiniCorr is ideally suited for spot-checking or bedside
monitoring in home care and clinical environments, as well
as during sleep screenings.
Features and Benefits
-
Reliably
measure SpO2, Pulse Rate and Pulse Strength.
-
SAC
Technology informs operator of excess artifacts and improper sensor
positioning.
-
Easy-to-use, 5 key operation.
-
Infrared
link to printer.
-
Lightweight for easy transport.
-
Battery
power or battery bypass.
Materials
Included
-
Adult
Finger Sensor
-
Manual
-
6
AA batteries Technical Specifications
Displays,
Indicators & Keys
-
SpO2 LED numeric display: 0.3 inch (7.62mm) high
-
Pulse Rate LED numeric display: 0.3 inch (7.62mm) high
-
Pulse Strength: Logarithmically scaled 8-segment LED bar graph
-
Low Battery LED: lights when approximately 30 minutes of battery life
remains
-
Audio Disabled LED: lights to indicate temporary or indefinite status
-
Sensor LED: lights to indicate an Oximeter sensor problem
-
Artifact LED: lights to indicate the presence of excess motion or other
artifact
-
Power Green LED: indicates monitor is connected to AC power
-
Keys: On/Off; Alarm Select, Alarm Silence; UP arrow - Alarm Limit/Pulse
Volume
Adjustment; DOWN arrow - Alarm Limit/ Pulse Volume Adjustment;
Print/Clear
SpO2: -
Range: 0-99% functional SpO2 (1% increments)
-
Accuracy: ±2% at 70-99%; ±3% at 50-69%
-
Averaging: 4, 8, or 16 pulse beat average
-
Alarm Range:
- >
High - 50%-99%
& OFF (1% steps)
- > Low - OFF & 50%-99% (1% steps)
Pulse Rate: -
Range: 30-254 bpm
-
Accuracy: ±2% or 2bpm, whichever greater at 30-254 bpm
-
Alarm Range:
- >
High - 5-250
bpm & OFF (5 bpm steps)
- > Low - OFF & 5-250 bpm (5
bpm steps)
- Averaging: 8 or 16 second average
Trend Storage -
Parameters: SpO2, Pulse Rate, Pulse Strength, Relative Time
-
Interval: 4 seconds to 30 seconds
-
Capacity: 12 hours of storage at 4 sec. interval
-
24 hours of storage at 8 sec. interval
-
90 hours of storage at 30 sec. interval
Power -
Battery 6 standard AA alkaline cells
-
Battery Life 24 hours without printing
-
AC Power (optional) External 12 VDC power supply
Printer Output: -
Data Log Mode: SpO2, and Pulse Rate printed every 5 seconds.
-
Trend Data Mode: Up to 90 hours of previously stored data, collected
from 1 to
99 patients.
-
PC Communications: RS232C With the oximeter sensor removed, trend data
can be transmitted
from the
sensor connector to PC serial port by way of a null modem cable.
Physical: -
Dimensions 6.6” H x 2.75” W x 1.43” D, 16.7 cm x 7.0 cm x 3.6 cm
- Weight:
16 oz.
(454 grams) with batteries
Environment: -
Operating Temperature: 32°F to 131°F (0°C to 55°C)
-
Storage Temperature: -40°F to 167°F (-40°C to 75°C)
-
Relative (non-condensing): Humidity 15-95% (storage), 10-95% (operating)
Standards: -
EN60601-1 (IEC 60601-1), UL544, CSA Std C22.2
-
No. 125-1984, EN60601-1-1, EMC: EN60601-1-2
|
|
|
|
|
|
|