| Oxygen Saturation Range |
0–100% SpO2 |
| Pulse Rate Range |
18–321 BPM |
|
Displays Pulse Strength: Alarm Indicator: Alarm Silenced: Numeric Displays: Low Battery:
|
LED, Bar graph, tri-color segments LED, bi-color LED, amber 3-digit, 7-segment LEDs, green LED, amber
|
Saturation Accuracy (Arms)a:
|
70–100% |
| NO MOTION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 2 digits |
± 3 digits |
| Disposable: 6000 Series |
± 2 digits |
± 3 digits |
| MOTION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 2 digits |
NA | |
Pulse Rate Accuracy:
|
18–300 BPM (no motion) 40–240 BPM (motion) 40–240 BPM (low perfusion) |
| NO MOTION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 3 digits |
± 3 digits |
| Disposable: 6000 Series |
± 3 digits |
± 3 digits |
| MOTION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 5 digits |
± 5 digits |
| LOW PERFUSION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 3 digits |
± 3 digits |
| Disposable: 6000 Series |
± 3 digits |
± 3 digits | |
Memory:
|
70 hours (assuming continuous operation) |
Temperature (Operating): Temperature (Storage/Transportation) |
0° to +40°C (32°F to 104°F) -30° C to +50°C (-2°F to 122°F) |
Humidity (Operating): Humidity (Storage/Transportation): |
10–90% noncondensing 10–95% noncondensing |
Altitude (Operating): Hyperbaric Pressure: |
Up to 12,000 meters (40,000 feet) Up to 4 atmospheres |
| Power Requirements (Mains): |
100–240 VAC 50–60 Hz |
|
Internal Power: Battery: Operating Life (fully charged battery): Storage Life: Recharge Rate:
|
7.2 volt NiMH battery pack 16 hours minimum 21 days minimum 4 hours maximum
|
| Dimensions: |
Approximately 219 mm (8.6”) W x 92 mm (3.6”) H x 142 mm (5.6”) D |
| Weight: |
Approximately 900 grams (2 lbs) with battery |
| Warranty: |
3 years |
| Classification per IEC 60601-1/CSA601.1/UL60601-1: |
| Type of Protection: |
Internally powered (on battery power) |
| Degree of Protection: |
Type BF-Applied Part |
| Mode of Operation: |
Continuous |
| Enclosure Degree of Ingress Protection: |
IPX2 |
|
Analog Outputs: SpO2 Output Range: Pulse Rate Output Range: Event Marker: Accuracy: Load Current:
|
0-1 VDC (0–100% SpO2), 1.27 VDC (out of track) 0-1 VDC (0–300 BPM), 1.27 VDC (out of track) 0 V (no event), 1 V (event occurred) ± 2% (SpO2), ± 5% (Pulse Rate) 2 mA maximum
|
a ± Arms represents approximately 68% of measurements
Specifications are subject to change without notice. |