Oxygen Saturation Range
|
0–100% SpO2
|
Pulse Rate Range
|
18–321 BPM
|
Displays Pulse Strength: Alarm Indicator: Alarm Silenced: Numeric Displays: Low Battery:
|
LED, Bar graph, tri-color segments LED, bi-color LED, amber 3-digit, 7-segment LEDs, green LED, amber
|
Saturation Accuracy (Arms)a:
|
70–100%
|
NO MOTION
|
Adults/Pediatrics
|
Neonates
|
Reusable: Finger Clip
|
± 2 digits
|
± 3 digits
|
Disposable: 6000 Series
|
± 2 digits
|
± 3 digits
|
MOTION
|
Adults/Pediatrics
|
Neonates
|
Reusable: Finger Clip
|
± 2 digits
|
NA
|
|
Pulse Rate Accuracy:
|
18–300 BPM (no motion) 40–240 BPM (motion) 40–240 BPM (low perfusion)
|
NO MOTION
|
Adults/Pediatrics
|
Neonates
|
Reusable: Finger Clip
|
± 3 digits
|
± 3 digits
|
Disposable: 6000 Series
|
± 3 digits
|
± 3 digits
|
MOTION
|
Adults/Pediatrics
|
Neonates
|
Reusable: Finger Clip
|
± 5 digits
|
± 5 digits
|
LOW PERFUSION
|
Adults/Pediatrics
|
Neonates
|
Reusable: Finger Clip
|
± 3 digits
|
± 3 digits
|
Disposable: 6000 Series
|
± 3 digits
|
± 3 digits
|
|
Memory:
|
70 hours (assuming continuous operation)
|
Temperature (Operating): Temperature (Storage/Transportation)
|
0° to +40°C (32°F to 104°F) -30° C to +50°C (-2°F to 122°F)
|
Humidity (Operating): Humidity (Storage/Transportation):
|
10–90% noncondensing 10–95% noncondensing
|
Altitude (Operating): Hyperbaric Pressure:
|
Up to 12,000 meters (40,000 feet) Up to 4 atmospheres
|
Power Requirements (Mains):
|
100–240 VAC 50–60 Hz
|
Internal Power: Battery: Operating Life (fully charged battery): Storage Life: Recharge Rate:
|
7.2 volt NiMH battery pack 16 hours minimum 21 days minimum 4 hours maximum
|
Dimensions:
|
Approximately 219 mm (8.6”) W x 92 mm (3.6”) H x 142 mm (5.6”) D
|
Weight:
|
Approximately 900 grams (2 lbs) with battery
|
Warranty:
|
3 years
|
Classification per IEC 60601-1/CSA601.1/UL60601-1:
|
Type of Protection:
|
Internally powered (on battery power)
|
Degree of Protection:
|
Type BF-Applied Part
|
Mode of Operation:
|
Continuous
|
Enclosure Degree of Ingress Protection:
|
IPX2
|
Analog Outputs: SpO2 Output Range: Pulse Rate Output Range: Event Marker: Accuracy: Load Current:
|
0-1 VDC (0–100% SpO2), 1.27 VDC (out of track) 0-1 VDC (0–300 BPM), 1.27 VDC (out of track) 0 V (no event), 1 V (event occurred) ± 2% (SpO2), ± 5% (Pulse Rate) 2 mA maximum
|
a ± Arms represents approximately 68% of measurements
Specifications are subject to change without notice.
|