CO2 Units: mmHg; Torr; %; kPa Method: Non-dispersive infrared; autocalibrating N2 O Compensation: Manual; On/Off Range: 0 to 99mmHg; 0 to 99 Torr; 0 to 12.5%; 0 to 12.5 kPa Resolution: 1mmHg; 1 Torr; 0.1%; 0.1 kPa Sample Rate: 200 ml/minAccuracy: +2mmHg, ±0.3 volume %, ±0.3 kPa, or ±2 Torr or ±4% of reading, whichever is greater, at 200ml/min and respiratory rate of <= 120bpm Rise Time: 170ms (10 to 90%) at 200 ml/min Invasive Blood Pressure Channels: 1 or 3 Transducer Sites: Selectable; ART, PA, LA, RA, LV, RV, CVP, ICP Pressure Range: -10mmHg to 300mmHg Transducer: Sensitivity 5µV/V/mmHg; Freq. Response 0-200 Hz; Impedance >300 ohms Accuracy: ±1mmHg or ±1%, whichever is greater Scale Range: -10 to 10, 0 to20, 0 to 30, 0 to 40, 0 to 60, 0 to 100, 0 to 150, 0 to 200, 0 to 300 Alarms Indication: Audible; visual (EN 60601-1-8) Silence: Yes; 2 min or permanent Trend Memory Tabular: 24 hr at 30 sec intervals Display: Tabular, Graphical Display Screen: 12.1" active color TFT Resolution: 800 x 600 pixels Waveforms: 6, maximumWaveform Types: ECG Leads I, II, III, aVR, aVL, aVF, V; CO2, Respiration; Pleth; IBP; breath-by-breath CO2; ST Histograms; Mini-Trends; OXY-CRG Languages: English, French, German, Italian, Portuguese, Spanish, Russian Controls Keys: 10; membrane-activated Rotary Knob: Push and rotate; 24 steps/turn System Outputs Com Ports: Two RS 232-compatible DB9 ports; USB (for software downloads only) Nurse Call: Contact switch; 3.5mm stereo phone jack, standard 14mm length Defib/Sync: BNC connector Video Port: Serial VGA compatibleDC Out: 5 VDC for optional wireless accessoryExternal Module Port: 8-PIN circular connector Recorder Type: Thermal line printer Paper Speed: 12.5, 25, 50mm/sec
Mechanical Weight: Base unit without batteries: 12 lb; 5.4 kg Size: Base unit: 12 in (H) x 14.5 in (W) x 11.5 in (D); 30.5 cm (H) x 36.5 cm (W) x 29.2 cm (D) Power Requirements Voltage: 100 to 240 VAC; 50/60 Hz Consumption: 35 W, typical Battery: 1 sealed lead acidBattery Life: 2 hours with no external modulesRecharge Time: 4 to 6 hours, typical Environmental Operating Temperature: 32° to 104° F; 0° to 40° C Storage Temperature: 23° to 122° F; -5° to 50° C Operating Humidity: 15% to 90%, noncondensing
Classification Types of Protection: Class I equipment Degree of Protection: Type CF defib-proof Safety: EN 60601-1